Q

Which camera is suitable for fluorescence applications?

A

For fluorescence applications, high-sensitivity sensor cameras are needed. If exposure time is only a few seconds, non-cooled cameras are sufficient. E3ISPM and E3CMOS cameras are very suitable. Users can select different models based on their sensitivity parameters. However, if the signal is very weak and requires long exposure, such as 3-5 seconds or more, cooled cameras are needed. When cooled cameras are set to low temperature (usually 0 degrees to 5 degrees is sufficient), during long exposure, dark current and temperature-induced noise will remain at very low levels, and images obtained from the camera can be very clean.